Nickel-alloy thermocouples
Thermocouple Type E
Type E (chromel–constantan) has a high output (68 µV/°C), which makes it well suited to cryogenic use. Additionally, it is non-magnetic. Wide range is −50 °C to +740 °C and narrow range is −110 °C to +140 °C.
Thermocouple Type J
Type J (iron–constantan) has a more restricted range (−40 °C to +750 °C) than type K, but higher sensitivity of about
50 µV/°C. The Curie point of the iron (770 °C) causes a smooth change in the characteristic, which determines the upper temperature limit.
Thermocouple Type K
Type K (chromel–alumel) is the most common general-purpose thermocouple with a sensitivity of approximately 41 µV/°C. It is inexpensive, and a wide variety of probes are available in its −200 °C to +1350 °C (−330 °F to +2460 °F) range.
Type K was specified at a time when metallurgy was less advanced than it is today, and consequently characteristics may vary considerably between samples. One of the constituent metals, nickel, is magnetic; a characteristic of thermocouples made with magnetic material is that they undergo a deviation in output when the material reaches its Curie point, which occurs for type K thermocouples at around 185 °C.
Thermocouple Type M
Type M (82%Ni/18%Mo–99.2%Ni/0.8%Co) are used in vacuum furnaces for the same reasons as with type C. Upper temperature is limited to 1400 °C. It is less commonly used than other types.
Thermocouple Type N
Type N (Nicrosil–Nisil) thermocouples are suitable for use between −270 °C and +1300 °C, owing to its stability and oxidation resistance. Sensitivity is about 39 µV/°C at 900 °C, slightly lower compared to type K.
Thermocouple Type T
Type T (copper–constantan) thermocouples are suited for measurements in the −200 to 350 °C range. Often used as a differential measurement, since only copper wire touches the probes. Since both conductors are non-magnetic, there is no Curie point and thus no abrupt change in characteristics.
Type-T thermocouples have a sensitivity of about 43 µV/°C. Note that copper has a much higher thermal conductivity than the alloys generally used in thermocouple constructions, and so it is necessary to exercise extra care with thermally anchoring type-T thermocouples.
Platinum/rhodium-alloy thermocouples
Types B, R, and S thermocouples use platinum or a platinum/rhodium alloy for each conductor. These are among the most stable thermocouples, but have lower sensitivity than other types, approximately 10 µV/°C. Type B, R, and S thermocouples are usually used only for high-temperature measurements due to their high cost and low sensitivity.
Thermocouple Type B
Type B (70%Pt/30%Rh–94%Pt/6%Rh) thermocouples are suited for use at up to 1800 °C. Type-B thermocouples produce the same output at 0 °C and 42 °C, limiting their use below about 50 °C. The emf function has a minimum around 21 °C, meaning that cold-junction compensation is easily performed, since the compensation voltage is essentially a constant for a reference at typical room temperatures.
Thermocouple Type R
Type R (87%Pt/13%Rh–Pt) thermocouples are used up to 1600 °C.
Thermocouple Type S
Type S (90%Pt/10%Rh–Pt) thermocouples, similar to type R, are used up to 1600 °C.
Tungsten/rhenium-alloy thermocouples
These thermocouples are well suited for measuring extremely high temperatures. Typical uses are hydrogen and inert atmospheres, as well as vacuum furnaces. They are not used in oxidizing environments at high temperatures because of embrittlement. A typical range is 0 to 2315 °C, which can be extended to 2760 °C in inert atmosphere and to 3000 °C for brief measurements.
Thermocouple Type C
(95%W/5%Re–74%W/26%Re)
Thermocouple Type D
(97%W/3%Re–75%W/25%Re)
Thermocouple Type G
(W–74%W/26%Re)
Chromel–gold/iron-alloy thermocouples
In these thermocouples (chromel–gold/iron alloy), the negative wire is gold with a small fraction (0.03–0.15 atom percent) of iron. The impure gold wire gives the thermocouple a high sensitivity at low temperatures (compared to other thermocouples at that temperature), whereas the chromel wire maintains the sensitivity near room temperature.
It can be used for cryogenic applications (1.2–300 K and even up to 600 K). Both the sensitivity and the temperature range depend on the iron concentration. The sensitivity is typically around 15 µV/K at low temperatures, and the lowest usable temperature varies between 1.2 and 4.2 K.
Thermocouple Type P
Type P (55%Pd/31%Pt/14%Au–65%Au/35%Pd) thermocouples give a thermoelectric voltage that mimics the type K over the range 500 °C to 1400 °C, however they are constructed purely of noble metals and so shows enhanced corrosion resistance. This combination is also known as Platinel II.
Platinum/molybdenum-alloy thermocouples
Thermocouples of platinum/molybdenum-alloy (95%Pt/5%Mo–99.9%Pt/0.1%Mo) are sometimes used in nuclear reactors, since they show a low drift from nuclear transmutation induced by neutron irradiation, compared to the platinum/rhodium-alloy types.
Iridium/rhodium alloy thermocouples
The use of two wires of iridium/rhodium alloys can provide a thermocouple that can be used up to about 2000 °C in inert atmospheres.
Pure noble-metal thermocouples Au–Pt, Pt–Pd
Thermocouples made from two different, high-purity noble metals can show high accuracy even when uncalibrated, as well as low levels of drift. Two combinations in use are gold–platinum and platinum–palladium. Their main limitations are the low melting points of the metals involved (1064 °C for gold and 1555 °C for palladium). These thermocouples tend to be more accurate than type S, and due to their economy and simplicity are even regarded as competitive alternatives to the platinum resistance thermometers that are normally used as standard thermometers.